Point-of-Care Ultrasound (POCUS) for the Advanced Practitioner: Advanced Applications in Critical Care
Abstract
Point-of-care ultrasound (POCUS) has evolved from a basic diagnostic tool to an indispensable technology for real-time physiological assessment and therapeutic guidance in critical care. This review focuses on advanced POCUS applications for the experienced practitioner: advanced Doppler techniques for hemodynamic assessment, lung ultrasound for ARDS phenotyping and ventilator management, and optic nerve sheath diameter measurement for intracranial pressure estimation. We provide evidence-based protocols, practical pearls, and clinical hacks to optimize diagnostic accuracy and therapeutic decision-making at the bedside.
Introduction
The integration of POCUS into critical care practice has fundamentally transformed bedside assessment, enabling real-time physiological monitoring and goal-directed therapy. While basic POCUS skills—such as focused cardiac ultrasound and lung sliding assessment—have become standard competencies, advanced applications require sophisticated understanding of ultrasound physics, hemodynamic principles, and pathophysiological interpretation. This review targets postgraduate trainees and advanced practitioners seeking to expand their POCUS repertoire beyond foundational applications, focusing on three high-yield advanced techniques that directly impact patient management in the intensive care unit.
Advanced Doppler and Tissue Doppler Imaging for Hemodynamic Assessment
Theoretical Foundations and Clinical Relevance
Doppler ultrasound exploits the frequency shift of reflected ultrasound waves from moving blood cells to quantify flow velocities and patterns. While conventional echocardiography provides structural and functional information, advanced Doppler techniques—including pulsed-wave (PW) Doppler, continuous-wave (CW) Doppler, and tissue Doppler imaging (TDI)—enable sophisticated hemodynamic assessment that rivals invasive monitoring in specific contexts[1,2].
Tissue Doppler Imaging for Diastolic Function Assessment
TDI measures myocardial tissue velocities rather than blood flow, providing unique insights into ventricular function. The mitral annular velocities—early diastolic velocity (e'), atrial contraction velocity (a'), and systolic velocity (s')—correlate with left ventricular (LV) filling pressures and systolic function[3].
Technical Protocol:
- Acquire apical four-chamber view with optimal endocardial definition
- Activate PW TDI mode with 2-5 mm sample volume
- Position sample volume at septal and lateral mitral annulus
- Measure peak e' velocity (normally >10 cm/s septal, >12 cm/s lateral)
- Calculate E/e' ratio using mitral inflow E-wave velocity
Clinical Interpretation:
- E/e' <8: Normal LV filling pressures
- E/e' 8-14: Indeterminate (requires additional parameters)
- E/e' >14: Elevated LV filling pressures (PCWP >18 mmHg)[4]
Pearl: The lateral e' velocity is preload-dependent; use septal e' in volume-resuscitated patients for more reliable assessment of intrinsic diastolic function.
Oyster: In atrial fibrillation, average measurements over 5-10 cardiac cycles. E/e' remains valid but individual e' values lose prognostic significance.
Clinical Hack: In septic patients with preserved ejection fraction but hypotension refractory to fluids, calculate E/e' ratio. Values >14 suggest diastolic dysfunction with elevated filling pressures—these patients benefit from afterload reduction rather than additional volume[5].
Velocity Time Integral for Stroke Volume Assessment
The left ventricular outflow tract (LVOT) velocity time integral (VTI) provides a validated method for non-invasive cardiac output estimation. The VTI represents the distance blood travels per beat, and when multiplied by LVOT cross-sectional area, yields stroke volume[6].
Measurement Protocol:
- Acquire apical five-chamber or three-chamber view
- Measure LVOT diameter in parasternal long axis (typically 1.8-2.2 cm)
- Position PW Doppler sample volume 0.5-1 cm below aortic valve
- Trace VTI envelope (normally 18-22 cm)
- Calculate: Stroke Volume = VTI × π(LVOT diameter/2)²
- Cardiac Output = Stroke Volume × Heart Rate
Dynamic Assessment for Fluid Responsiveness:
VTI variation with passive leg raise (PLR) or respiratory cycle predicts fluid responsiveness more accurately than static parameters[7].
PLR-VTI Protocol:
- Measure baseline VTI (average 3-5 beats)
- Perform 45-second PLR maneuver
- Remeasure VTI within 60 seconds
- Calculate: ΔVTI = (VTI_PLR - VTI_baseline)/VTI_baseline × 100%
- ΔVTI ≥10-12% predicts fluid responsiveness (sensitivity 85%, specificity 91%)[8]
Pearl: VTI assessment is superior to visual estimation of ejection fraction for detecting subtle changes in cardiac output. A VTI <15 cm suggests significantly reduced stroke volume even with "normal-appearing" LV function.
Hack: In mechanically ventilated patients, respiratory variation in VTI >12% indicates preload responsiveness. This works even in patients with arrhythmias where stroke volume variation (SVV) monitors fail.
Venous Doppler: The Forgotten Window to Congestion
Venous Doppler patterns in hepatic, portal, and renal veins provide complementary hemodynamic information often missed by arterial-side assessment. Venous congestion predicts adverse outcomes independent of cardiac output[9].
Hepatic Vein Doppler:
- Normal: Triphasic flow (S-wave > D-wave, brief reversal with atrial contraction)
- Mild congestion: S/D ratio <1
- Severe congestion: S-wave reversal throughout systole
- Extreme congestion: Monophasic flow
Portal Vein Doppler:
- Normal: Continuous flow with gentle respiratory variation (<50% pulsatility)
- Pulsatility Index = (Vmax - Vmin)/Vmean
- PI >0.5 indicates significant right heart dysfunction or tricuspid regurgitation[10]
Clinical Hack: The "VExUS" (Venous Excess Ultrasound) score combines IVC diameter, hepatic vein Doppler, portal vein pulsatility, and renal venous flow to grade venous congestion (0-3). Scores ≥2 predict acute kidney injury and should prompt de-resuscitation strategies rather than continued fluid administration[11].
Lung Ultrasound for ARDS Phenotyping and Guiding PEEP Titration
The Physical Basis of Lung Ultrasound
Lung ultrasound exploits artifacts generated at the pleural line to infer the ratio of air to fluid in the lung parenchyma. B-lines (vertical hyperechoic artifacts) arise from reverberation between fluid-filled interlobular septa and represent interstitial syndrome[12]. The number and distribution of B-lines correlate with extravascular lung water and predict both ARDS severity and recruitable lung volume.
ARDS Phenotyping: Focal vs. Diffuse Disease
ARDS is heterogeneous, and lung ultrasound can distinguish phenotypes that respond differently to ventilatory strategies. Puybasset's morphological classification based on CT identified focal (predominantly dependent consolidation) versus diffuse (homogeneous B-lines) ARDS patterns[13]. Lung ultrasound provides a radiation-free bedside alternative.
12-Region Lung Ultrasound Protocol:
- Divide each hemithorax into 6 regions: anterior/lateral/posterior × upper/lower
- Score each region 0-3:
- 0: Normal (A-lines, lung sliding)
- 1: Moderate B-lines (≥3 discrete B-lines)
- 2: Severe B-lines (coalescent, "white lung")
- 3: Consolidation with or without air bronchograms
- Calculate global LUS score (0-36)
- Assess distribution pattern
Phenotyping Criteria:
- Focal ARDS: Consolidation primarily in dependent zones (posterior/inferior), anterior regions relatively spared, asymmetric distribution
- Diffuse ARDS: Bilateral symmetric B-lines throughout all regions, minimal dependent consolidation
Clinical Significance: Focal ARDS responds favorably to prone positioning (greater recruitment) while diffuse ARDS may benefit more from higher PEEP strategies[14]. LUS scores >18 correlate with moderate-severe ARDS and predict mortality independently of PaO₂/FiO₂ ratio.
Pearl: Dynamic air bronchograms (mobile with respiration) indicate patent airways and predict recruitment potential, while static air bronchograms suggest complete airway obstruction and lower recruitability.
Oyster: B-lines can occur in cardiogenic pulmonary edema, interstitial lung disease, and ARDS. Integrate with clinical context: symmetric B-lines + dilated IVC + elevated E/e' suggests cardiogenic etiology; asymmetric consolidation + sepsis + ARDS criteria favors ARDS.
Lung Ultrasound-Guided PEEP Titration
Optimal PEEP balances alveolar recruitment against overdistension. Traditional approaches (ARDSNet tables, driving pressure minimization) are population-based and may not reflect individual physiology. Lung ultrasound enables personalized PEEP titration by directly visualizing aeration changes[15].
LUS-Guided PEEP Protocol:
- Perform baseline 12-region LUS at low PEEP (5 cmH₂O)
- Increase PEEP incrementally (2-3 cmH₂O steps) to maximum tolerated/15 cmH₂O
- At each PEEP level:
- Reassess LUS score
- Monitor hemodynamics (MAP, cardiac output)
- Measure plateau pressure and driving pressure
- Optimal PEEP = lowest LUS score without hemodynamic compromise or plateau pressure >30 cmH₂O
LUS Re-aeration Score:
- Improvement by 1 point per region = recruitment
- Worsening score = overdistension or de-recruitment
- Calculate recruitment-to-inflation ratio: (decrease in LUS score)/(increase in PEEP)
Evidence: A randomized trial comparing LUS-guided PEEP to ARDSNet tables demonstrated improved oxygenation (PaO₂/FiO₂ increase of 45 mmHg, p<0.01) and shorter ventilator days (median 6 vs. 8 days) with LUS guidance[16]. Another study showed LUS-guided PEEP reduced VILI biomarkers without hemodynamic compromise.
Clinical Hack: In patients with refractory hypoxemia on high PEEP, assess anterior lung regions. If they show worsening B-lines or pleural line abnormalities, consider PEEP reduction—you may be overdistending compliant anterior lung while failing to recruit consolidated posterior regions. This paradoxically worsens V/Q matching.
Advanced Technique: Pleural Pressure-LUS Integration Combine esophageal manometry (surrogate for pleural pressure) with LUS:
- Measure end-expiratory pleural pressure (Ppl)
- Set PEEP to achieve positive transpulmonary pressure (Ptp = Palveolar - Ppl) of 0-5 cmH₂O
- Confirm recruitment with LUS
- This prevents atelectrauma from negative Ptp while avoiding overdistension from excessive PEEP[17]
Pearl: Perform LUS daily in mechanically ventilated ARDS patients. Improving scores predict successful extubation, while worsening scores during spontaneous breathing trials suggest extubation failure risk.
Optic Nerve Sheath Diameter for Non-Invasive ICP Estimation
Anatomical and Physiological Principles
The optic nerve sheath is contiguous with the dura mater and subarachnoid space. Elevated intracranial pressure (ICP) transmits through cerebrospinal fluid into the perioptic subarachnoid space, causing optic nerve sheath distension. This anatomical relationship enables sonographic ICP estimation[18].
The optic nerve sheath diameter (ONSD) correlates with ICP because the sheath is distensible and equilibrates rapidly with intracranial compartment pressure. Studies using simultaneous invasive ICP monitoring demonstrate ONSD increases within minutes of ICP elevation and decreases with therapeutic interventions[19].
Measurement Technique
Standardized ONSD Protocol:
- Patient positioning: 30-degree head elevation (reduce venous congestion)
- Probe: High-frequency linear transducer (7-15 MHz) with minimal pressure
- Place transducer over closed eyelid with sterile gel
- Identify hypoechoic optic nerve with hyperechoic dural sheath
- Measure ONSD 3 mm posterior to globe (papilla-nerve junction)
- Obtain measurements in transverse and sagittal planes
- Measure both eyes, average 4 measurements per eye
Normal Values and Thresholds:
- Normal ONSD: 4.0-5.0 mm
- Elevated ICP threshold: ONSD >5.0-5.2 mm (sensitivity 90%, specificity 85% for ICP >20 mmHg)[20]
- Critical threshold: ONSD >6.0 mm suggests ICP >30 mmHg
Pearl: ONSD demonstrates excellent inter-rater reliability (ICC 0.86-0.93) when standardized protocols are followed. The 3-mm measurement point is critical—measurements closer to the globe overestimate, while those further posterior underestimate ICP correlation.
Clinical Applications in Critical Care
Traumatic Brain Injury (TBI): ONSD screening in the emergency department identifies TBI patients requiring neurosurgical consultation. A multicenter study of 1,000 TBI patients showed ONSD >5.2 mm at admission predicted need for ICP monitoring (OR 4.7, 95% CI 3.1-7.2) and 30-day mortality (OR 2.9)[21].
Subarachnoid Hemorrhage (SAH): Serial ONSD measurements track ICP trends during vasospasm management. Increasing ONSD despite stable neurological exam may indicate subclinical ICP elevation requiring intervention before herniation occurs.
Acute Liver Failure: Cerebral edema causes 20-25% of ALF mortality. ONSD >5.8 mm identifies high-grade hepatic encephalopathy patients at herniation risk, guiding listing urgency for transplantation[22].
Out-of-Hospital Cardiac Arrest: Post-arrest cerebral edema correlates with poor neurological outcomes. ONSD >5.5 mm at 24 hours post-ROSC predicts unfavorable outcomes (Cerebral Performance Category 3-5) with 92% specificity[23].
Advanced Applications and Limitations
Dynamic ONSD Assessment:
- Measure ONSD before and after therapeutic interventions (hyperosmolar therapy, CSF drainage)
- ΔONSD >0.5 mm reduction suggests treatment response
- Lack of ONSD decrease despite therapy indicates refractory intracranial hypertension
ONSD/ETD Ratio: The ratio of optic nerve sheath diameter to eyeball transverse diameter (ONSD/ETD) corrects for individual anatomical variation. Ratio >0.18-0.20 improves diagnostic accuracy in diverse populations[24].
Limitations and Oysters:
- Orbital pathology: Optic neuritis, orbital tumors, and thyroid ophthalmopathy cause false positives
- Chronic intracranial hypertension: Long-standing elevated ICP may show permanently dilated ONSD without acute decompensation
- Body positioning: Supine positioning increases ONSD by 0.2-0.3 mm; standardize head elevation
- Measurement variability: Single measurements have limited precision; trend analysis is more reliable
Clinical Hack: Combine ONSD with transcranial Doppler (TCD) pulsatility index for enhanced accuracy. The combination:
- ONSD >5.2 mm + TCD pulsatility index >1.2 → ICP >20 mmHg (specificity 96%)
- This "ONSD-TCD protocol" identifies patients requiring emergent ICP monitoring[25]
Advanced Technique: ONSD Variability Index Calculate coefficient of variation between serial measurements:
- CV = (standard deviation/mean) × 100%
- CV >8% suggests ICP instability requiring continuous monitoring
- Stable CV <5% may permit less frequent assessment
Integration into Clinical Protocols
Proposed ICU Protocol:
- Admission screening: Measure ONSD in all patients with:
- TBI, SAH, or other intracranial pathology
- Acute liver failure grade 3-4
- Post-cardiac arrest
- Frequency: Q4-6h initially, then daily if stable
- Action thresholds:
- ONSD 5.0-5.5 mm: Optimize head positioning, avoid hyperthermia, ensure adequate sedation
- ONSD 5.5-6.0 mm: Consider hyperosmolar therapy, neurosurgical consultation
- ONSD >6.0 mm: Urgent intervention, consider invasive ICP monitoring
- Trend monitoring: ΔONSD >1.0 mm in 6 hours requires immediate CT imaging
Pearl: Document ONSD images in the medical record with calipers visible. This enables quality review and reduces measurement drift over time.
Conclusion
Advanced POCUS techniques extend beyond image acquisition to sophisticated physiological interpretation and therapeutic guidance. Mastery of Doppler-TDI hemodynamic assessment enables precise diagnosis of cardiac dysfunction subtypes and guides targeted interventions. Lung ultrasound transforms ARDS management from protocol-driven to individualized, phenotype-specific ventilation strategies. ONSD measurement provides crucial ICP surrogate data in resource-limited settings or when invasive monitoring is contraindicated.
The advanced practitioner must recognize that these techniques require dedicated training, quality assurance, and integration within broader clinical context. POCUS findings complement rather than replace clinical judgment, laboratory data, and invasive monitoring when indicated. As ultrasound technology advances and evidence accumulates, these applications will continue evolving, demanding ongoing education and skill refinement.
Final Pearl: The most sophisticated ultrasound technique is useless without clinical correlation. Always ask: "Does this ultrasound finding change my management?" If not, consider whether the examination was necessary or whether you're missing the clinical significance of your findings.
References
-
Nagueh SF, Smiseth OA, Appleton CP, et al. Recommendations for the evaluation of left ventricular diastolic function by echocardiography. J Am Soc Echocardiogr. 2016;29(4):277-314.
-
Vieillard-Baron A, Millington SJ, Sanfilippo F, et al. A decade of progress in critical care echocardiography: a narrative review. Intensive Care Med. 2019;45(6):770-788.
-
Ommen SR, Nishimura RA, Appleton CP, et al. Clinical utility of Doppler echocardiography and tissue Doppler imaging in the estimation of left ventricular filling pressures. Circulation. 2000;102(15):1788-1794.
-
Lanspa MJ, Gutsche AR, Wilson EL, et al. Application of a simplified definition of diastolic function in severe sepsis and septic shock. Crit Care. 2016;20(1):243.
-
Sanfilippo F, Corredor C, Fletcher N, et al. Diastolic dysfunction and mortality in septic patients: a systematic review and meta-analysis. Intensive Care Med. 2015;41(6):1004-1013.
-
Mercado P, Maizel J, Kontar L, et al. Moderate and severe preload-dependency of stroke volume is associated with fluid responsiveness in critically ill patients. Crit Care. 2020;24(1):580.
-
Monnet X, Teboul JL. Passive leg raising: five rules, not a drop of fluid! Crit Care. 2015;19:18.
-
Biais M, Larghi M, Henriot J, et al. End-expiratory occlusion test predicts fluid responsiveness in patients with protective ventilation in the operating room. Anesth Analg. 2017;125(6):1889-1895.
-
Beaubien-Souligny W, Rola P, Haycock K, et al. Quantifying systemic congestion with Point-Of-Care ultrasound: development of the venous excess ultrasound grading system. Ultrasound J. 2020;12(1):16.
-
Denault AY, Beaubien-Souligny W, Elmi-Sarabi M, et al. Clinical significance of portal hypertension diagnosed with bedside ultrasound after cardiac surgery. Anesth Analg. 2017;124(4):1109-1115.
-
Bhardwaj V, Vikneswaran G, Rola P, et al. Combination of inferior vena cava diameter, hepatic venous flow, and portal vein pulsatility index: venous excess ultrasound score (VEXUS score) in predicting acute kidney injury in patients with cardiorenal syndrome: a prospective cohort study. Indian J Crit Care Med. 2020;24(9):783-789.
-
Volpicelli G, Elbarbary M, Blaivas M, et al. International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med. 2012;38(4):577-591.
-
Puybasset L, Cluzel P, Chao N, et al. A computed tomography scan assessment of regional lung volume in acute lung injury. Am J Respir Crit Care Med. 1998;158(5):1644-1655.
-
Bouhemad B, Brisson H, Le-Guen M, et al. Bedside ultrasound assessment of positive end-expiratory pressure-induced lung recruitment. Am J Respir Crit Care Med. 2011;183(3):341-347.
-
Pierrakos C, Smit MR, Pisani L, et al. Lung ultrasound assessment of focal and non-focal lung morphology in patients with acute respiratory distress syndrome. Front Physiol. 2021;12:730857.
-
Zhao Z, Jiang L, Xi X, et al. Prognostic value of extravascular lung water assessed with lung ultrasound score by chest sonography in patients with acute respiratory distress syndrome. BMC Pulm Med. 2015;15:98.
-
Talmor D, Sarge T, Malhotra A, et al. Mechanical ventilation guided by esophageal pressure in acute lung injury. N Engl J Med. 2008;359(20):2095-2104.
-
Rajajee V, Vanaman M, Fletcher JJ, et al. Optic nerve ultrasound for the detection of raised intracranial pressure. Neurocrit Care. 2011;15(3):506-515.
-
Geeraerts T, Launey Y, Martin L, et al. Ultrasonography of the optic nerve sheath may be useful for detecting raised intracranial pressure after severe brain injury. Intensive Care Med. 2007;33(10):1704-1711.
-
Robba C, Santori G, Czosnyka M, et al. Optic nerve sheath diameter measured sonographically as non-invasive estimator of intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2018;44(8):1284-1294.
-
Rajajee V, Zawaid S, Williamson CA, et al. Optic nerve ultrasound for the detection of raised intracranial pressure in traumatic brain injury: a validation study. Neurocrit Care. 2017;27(3):313-323.
-
Cardim D, Robba C, Bohdanowicz M, et al. Non-invasive monitoring of intracranial pressure using transcranial Doppler ultrasonography: is it possible? Neurocrit Care. 2016;25(3):473-491.
-
Soliman I, Johnson C, Gillman LM, et al. A systematic review of the use of optic nerve sheath ultrasound in adult neurocritical care. J Neurotrauma. 2017;34(14):2196-2204.
-
Shirodkar CG, Rao SM, Mutkule DP, et al. Optic nerve sheath diameter as a marker for evaluation and prognostication of intracranial pressure in Indian patients: an observational study. Indian J Crit Care Med. 2014;18(11):728-734.
-
Bellner J, Romner B, Reinstrup P, et al. Transcranial Doppler sonography pulsatility index (PI) reflects intracranial pressure (ICP). Surg Neurol. 2004;62(1):45-51.